Below are instructions on how to operate Conviron Central Management System for the following growth chambers and environmental rooms:
- 2016C, 2016D, 2016E, 2033-03ER, 2033-04ER, 3092-01GC, 3092-02GC, 3092-03GC, 3092-04GC
- 3028-01ER, 3028-06GC, 3028-07GC, 3028-08GC and 3028-09GC
These instructions do not replace the manufacturer's manual for Central Management System [available here].
View or download environmental data log
- On Conviron CM desktop, click on Central Management icon.
- Right-click on a chamber name. Click on Trend graph.
- Change entry under LLC Date, Days, and Plot accordingly.
Create or edit an environment setting
- Open Central Management.
- Right-click on a chamber's name. Click on Virtual Chamber.
- Create and edit program
- Under Main Menu > Programs, click on the create or edit icon.
- Type or select a Program ID, then click on Select.
- Create your first line: Click on Insert line. Type 00:00 under time. Indicate desired Temperature (°C), Light intensity (0 to 750 µMOL), and Light Level (0,1,2)
- Click on Insert line again to create your second line. Input the end time for first line. All other input should be the same as your first line.
- Click on Insert line for the third time. Indicate desired environmental settings.
- Repeat Step 3.4. Note that all parameters will ramp between Line 2 and Line 3 settings.
- Repeat Steps 3.5 and 3.6, if necessary. Note that the last line should indicate 23:59 under Time.
- Click Upload program into chamber. Confirm by clicking Yes. Then click on Ok, then Close.
- Create and edit schedule
- Select Edit on Schedule.
- Under Program number, type the Program ID number that you wish to run.
- On the cell on the right, indicate the number of cycles you wish to run the program.
- Click insert entry to add more programs.
- Click on Upload schedule into chamber
- Confirm with Yes. Click Ok.
- Modify the rate at which temperature changes between settings
- Click on Ramp/Step, then click on Change
- Type the Speed at which temperature should ramp between settings (Min/°C). It is currently set to 12.
- Click Start/stop on Main Menu. Click Ok.
- Click Start chamber from the selected row
- Remember to also manually turn on the lights inside the chamber.
Add users to receive an alarm:
- On desktop, click on the Central Management icon
- On the left pane, click on Edit Users.
- On a new window, click on Add New and then add Username and Email. Click on Ok.
- Go to Step 5 or Step 6.
- To receive an alarm for all chambers in CM:
- On the menu bar, select Tools>Network subscribers.
- On a new window, click on the bar next to the asterisk. Choose the name of the user you wish to add.
- Click Ok.
- To receive alarms from specific chambers only:
- On the main window, right-click on the chamber of interest. Then click on Properties.
- On the new window, select the Subscribers tab.
- Click on the bar opposite the asterisk symbol.
- Choose the name of the user you wish to add.
- Click Ok.
You may also remote to Conviron CM desktop:
Create a service request through ServiceDesk to ask assistance in setting up access to Conviron CM desktop in DV2033. Ask for the following to be done for you:
- On your computer:
- Install UTMVPN.
- Install Remote Desktop app. Add remote connection to Conviron CM desktop.
- On Conviron CM desktop in DV2033:
- Assist you in first time log-in using your utorid and password.
- Add desktop shortcut for C:\Program Files (x86)\CM\CentralManagement.
Add utm.growth.facilities on the same service request for staff to authorize your access.
Published 2020-10-21, last updated 2024-09-30